Connect the Go Direct® Chloride Ion-Selective Electrode BNC to our Go Direct® Ion-Selective Electrode Amplifier to measure the concentration of chloride (Cl–) ions in aqueous samples.
The Go Direct® Vernier Chloride Ion-Selective Electrode BNC is a BNC-terminated electrode used to measure the concentration of chloride (Cl–) ions in aqueous samples. Requires a Go Direct® Ion-Selective Electrode Amplifier.
Specifications
Range (mV):
Go Direct® Ion-Selective Electrode Amplifier: -1000 mV to +1000 mV
Range (concentration): 1 to 35,000 mg/L (or ppm)
Reproducibility (precision): ±30mV
Interfering ions: CN–, Br–, I–, S2-, OH–, NH3
pH range: 2–12 (no pH compensation)
Temperature range: 0–80°C (no temperature compensation)
Electrode slope: 56 ±3 mV/decade at 25°C
Electrode resistance: 1 to 5 MΩ
Minimum sample size: must be submerged 2.8 cm (1.1 in)